Ensure all plasma samples are centrifuged according to standard laboratory protocol (e.g., 15 minutes at 2500g to obtain platelet-poor plasma).
: Only authorized personnel should perform adjustments with the cover open.
Ignoring these steps can lead to inconsistent reagent dispensing and volumetric errors. diagnostica stago start 4 coagulation analyzer manual
Provides information about symbols used in the analyzer, including safety warnings and disposal guidelines. Start 4 PDF - Scribd
: As the technician adds a reagent, a timer starts. The magnetic field causes the metal ball to oscillate back and forth.
This guide is intended for informational purposes. Always refer to the official user manual provided by Diagnostica Stago for the most current and detailed safety information. For official support or to request a replacement
Intended for service engineers, this section details the installation requirements: floor space, vibration dampening, line voltage (110V/220V), and network connectivity for Stago’s middleware (if applicable).
When the Stago Start 4 encounters an issue, it displays specific flags or error messages. Below are the most common issues and how to resolve them. Error / Symptom Potential Cause Corrective Action
Grounded electrical outlet matching the voltage specified on the instrument's serial plate (typically 110/220V AC, 50/60 Hz). Initial Power-On Sequence : Only authorized personnel should perform adjustments with
Results marked with asterisks (*) often indicate that the reagent was not added quickly enough or the pipette was not used correctly, requiring a rerun .
Dispense one metal ball per cuvette using the .
Requires specialized cuvette strips, magnetic stir-bars, and stainless steel balls for operation.