What's happening?

Note: Calibration should only be performed using specialized non-inductive test loads and a high-frequency oscilloscope or specialized ESU analyzer.

Do not spray liquids directly onto the housing slots; use a dampened cloth to avoid internal dripping onto the high-voltage transformers. Reassembly Checklist

The service and instruction manuals for the Ellman Surgitron 4.0 Dual RF

A: No. The 4.0 Dual RF has a second frequency (1.7 MHz) and different output relay logic. The 2.0 manual lacks the bipolar/dual-frequency calibration procedures.

Internal circuitry contains potentials exceeding 1000V peak-to-peak during RF generation. Do not probe the mainboard blindly.

The Ellman Surgitron 4.0 Dual RF Go to product viewer dialog for this item.

Check the AC outlet power. Inspect and replace the rear-panel fuses with the exact rated replacements.

) and high-frequency leakage current matching standard IEC 60601-2-2 guidelines. Calibration and Waveform Verification

Pro tip: When searching, include the serial number range. Early 4.0 units (SN < 5000) used a different RF output board than later units (SN >= 5000). The correct service manual will list both schematics.